LabRAM Odyssey

LabRAM Odyssey basic configuration includes :

•    Silver metallic frame  (50th Anniversary Special Edition, 50 units numeroted)
•    800 mm focal length achromatic spectrometer
•    Open Space Microscope with Koelher illumination in Reflection
•    1 high brightness laser: 532 or 785 nm, 100 mW, TEM00
•    1 Ultra Low Frequency filter: cut-off down to 5 cm-1
•    Duoscan achromatic scanning without moving
•    SWIFT XS ultra-fast imaging down to 750 µs/spectrum
•    FIUV high linearity SynapseEM EMCCD
•    High resolution holographic gratings : 1800 & 3000 g/mm

•    Fast imaging low density gratings: 300 & 600 g/mm
•    XYZ motorized stage with encoders
•    Labspec6 64bits including following LabStore apps:
    -    EasyNavTM navigation and autofocus software
    -    ParticleFinderTM particle analysis software
    -    MVATM built-in Multivariate Analysis
    -    KIA HORIBA edition spectral library
•    SEM and AFM-Raman ready*
•    Lifetime application training support**

Manufactured by HORIBA Scientific

Specifications1

Unit

DUV @ 266nm 3000g/mm

VIS @ 532 nm 3000 g/mm

NIR @ 785 nm 1800 g/mm

Spectral Dispersion 

cm-1 / pix

< 0.83

< 0.13

< 0.12

Spatial resolution XY

µm

0.8 / 0.5

74XUVI NA 0.65

0.5 / 0.25

100X NA 0.95

0.8 / 0,5

100X NA 0.95

Confocal depth resolution (typical / best)

µm

 

1.5 / 0.8

100X NA 0.95

2 / 1.5

100X NA 0.95

PhotoLuminescence Spectral range2

nm

220 - 2100 (achromatic, no change of optics required)

High wavenumber cut-off3

cm-1

Up to 30 000

Low wavenumber cut-off3

cm-1

Down to 5

Imaging speed

ms/sp

Down to 0.75

Deep cooled Spectroscopic Multichannel Detector

Front Illuminated UV coated EMCCD

1600 x 200 pixel 16 x16µm² pixel size

Dark Current at -60°C

e-/pixel/s

< 0.002

Dimensions (Width x Depth x Height)

mm

1300 x 1194 x 473

1 Indicative values with 3 typical Raman lasers, other wavelengths available. Due to range of options and configurations of the system, this information is given as a guide to ultimate performance 
Acceptance criteria for a given configuration are available upon request.
2 Standard range with EMCCD is 300-1050nm. Requires optional InGaAs array detector to cover 1050-2100nm range
3 Laser wavelength dependant

PTI - Horiba

PT. HORIBA Indonesia

Ruko Jalur Sutera Jl. Jalur Sutera Blok 20 A No. 16-17 Tangerang, Banten 15144, Indonesia

Tel: +62 (21) 30448525
Fax: +62 (21) 30448521
Email: info-sci.id@horiba.com